The OEM-Flow A module has been designed for measuring the breathing gas flow in respiratory and anaesthesia machines. It works with state-of-the-art hot-wire flow sensors.
The flow sensor is controlled with a precision driving circuit.
The module is small-sized and works in a wide range of supply voltages.
An RS232 interface is used to interchange the data with the host system.
The board has an ADC resolution of 10 bit and a sampling rate of up to 200 samples/sec.
REF: 1030112001
Contact us for further information
Measuring principle |
Constant temperature mode |
Probe input |
1 x Hot-wire flow sensor |
Measuring range |
According to the sensor specification |
Accuracy |
+/- 5% in the range of 10 - 100 l/min (board only) |
Supply voltage |
5V DC; other voltages on request |
Current consumption |
Depending on gas flow |
Data interface |
RS232 |
Dimension |
Approximately 38 x 41 x 15mm |
Weight |
Approximately 10g |